Since the microscope is the primary instrument for data acquisition in spatial biology, it is useful to understand its parts and functionality to ensure the best data quality and to troubleshoot when there are issues with acquisition.

Fluorescence microscopes

Bright field microscopes

Troubleshooting

Blurry images

Blurry images are caused by two main things:

In both cases, the image will need to be reacquired, and the focal plane should be adjusted until the objects are in focus. Debris should be carefully removed from the optical path, either by gentle washing if it’s on the sample or with lens wipes if the debris is on the microscope.

Grainy images

In order for an object to be detected clearly, the real signal must be far enough above the noise (high signal-to-noise ratio) that the noise can be easily taken out through thresholding. Grainy images are characteristic of having low signal-to-noise ratio (SNR).

To improve noisy image quality, there are two options: increase signal strength or decrease noise. However, it is generally easier to increase signal strength because most noise sources are dictated by the physics of the microscope/camera/computer system.

Increasing signal strength

Decreasing noise